Installation and Operation Manual – AQUATEC 1000
System Description
Doc.-ID: AQUA_16012020_EN
7
1.6
Disruption of the Process
The analyzer system has to be kept in operation also in the event of the process being disrupted or if the plant is
powered off temporarily (e.g. at night or during the weekend). Frequently cooling down and heating up of the probe
results in thermal stress of the hot probe parts (heater, thermocouple and sensor) and reduces their product life.
ENOTEC will not accept any responsibility for resultant damage.
1.7
Storage instructions
ENOTEC equipment is to be stored in a dry and ventilated environment. Paint fumes, silicone sprays, etc. must be
avoided in the storage environment.
1.8
Name Plates
The name plate contains information about the line voltage, the nominated current, frequency, protection class, year
of manufacture, serial number, order number and system order code.
The system order code refers to information which is detailed in the system test report and supplied with the system.
SME-53 Electronic Unit
Probe Connection Box
Figure 3 -
Position of the name plates