Installation and Operation Manual – AQUATEC 1000
Service and Maintenance
Doc.-ID: AQUA_16012020_EN
29
5
Service and Maintenance
5.1
Exchange fuses
Warning
De-energize the system first!
Abb. 27
Position of the fuses
fuse
system voltage
ampere
nominal voltage
characteristic
Size
F3
115 / 230 V AC
0.5 A
250 V AC
T / L
5x20 mm
F4
115 / 230 V AC
0.5 A
250 V AC
T / L
5x20 mm
F5
115 / 230 V AC
1.0 A
250 V AC
M / L
5x20 mm
F1
230 V AC
2.0 A
250 V AC
T / H
5x20 mm
F1
115 V AC
4.0 A
250 V AC
T / H
5x20 mm
F2
115 / 230 V AC
4.0 A
250 V AC
M / H
5x20 mm
5.2
Flow rates for Test Air and Reference Air
The systems are factory-set to the correct amounts of test air and/or reference air. The instrument air versions are
designed for an inlet pressure of 1-10 bar. With a higher inlet pressure of 6 bar, it is necessary to readjust the flow of
reference air and/or test air.
The flow rates for air should be in the following range:
Test air
:
150l/h - 180l/h
Reference air
:
30l/h - 40l/h