13
Status Screens
(con’t)
g.
Press “Home” to return to Start Cycle screen
Lamps are designed for 12,000 run hours after which point, they will only provide 80% of
original UVC dosage output and will continue to degrade. The device will alert for lamp
replacement at 8,000 run hours. It is recommended that lamps be replaced at this time for
optimal dosage output.
2. Replace Lamp(s)
a. Press
“Service” button on Home screen
b. Lamp Replacement indicators will be
illuminated in Red
c. Replace lamps as indicated
d.
Press “Factory Settings” button
e. Enter Passcode (1470) and press enter
f.
Press “Reset Service Light” to clear alert
g.
Press “Home” to return to Start Cycle screen
Alarm Indicators
In the “Alarm” indicator is illuminated on the Home screen:
1.
Press “Alarm” indicator to observe the Alarm Log
2. Ensure all system alerts have been properly
addressed
3. Select HOME button to resume normal system
operation
4. Should Alarm persist, contact MaxAssure, Inc customer service (239) 643-2778