11
Status Screens
Cycle stop displays for normal system operation are covered on this page.
Emergency Stop
Should cycle deactivation occur by engagement of the Emergency Stop button, the following
method will be required to reset the system:
1. Observe Emergency Stop Active status on control screen
2. Release Emergency Stop by pulling up on the red knob
3.
Press “Cycle Status” to see treatment time remaining
4. Select
“Home” button to return to main screen
5. Press
“Start Cycle” to resume normal operation of system
PIR Sensor Cycle Interruption
Should cycle interruption occur by means of a PIR sensor activation, the following will be
required:
1. Observe the interruption status type on the control panel
2. Ensure no one is in treatment area
3. Select HOME button to return to main screen
4. Press START CYCLE to restart normal safe operation of system