Operation
21
5.3.1
Main menu
Access the main menu by holding down the
H
key until “AUTO MEASURING” is displayed.
For the main menu options and information about them, please see the following table.
5.3.2
AUTO MEASURING
The actions "calibration", "measuring" and "flushing" are triggered by time-control.
The settings for these actions are made in the "PARAMETER ENTRY" menu.
The respective action is displayed in the LC display. The most recently registered concentration
value is displayed until the end of the next measurement.
Otherwise, "wait" is displayed when
• the time of the first measurement has not yet been reached or
• the measuring interval has not yet expired.
!
Note!
"Measuring" flashes when the analyser is ready for the next measurement but has not yet received
the enable signal from the sample collector or the sample conditioning unit.
Selection
Display
Info
AUTO MEASURING
Calibration, measurement, flushing time-controlled
actions
PARAMETER ENTRY
Default settings for measuring ranges, alarm values,
calibration, flushing
CONFIGURATION
Basic settings such as parameters, measuring units,
arrangement of analog outputs and alarm values (NO,
NC), date, time, offset values
LANGUAGE
Selecting menu language
ERROR DISPLAY
Displaying error messages
SERVICE
Manually switching valves and pumps
DATA MEMORY 1
Last 1024 measured values channel 1
DATA MEMORY 2
(Two-channel version only)
Last 1024 measured values channel 2
AUTO MEASURING
PARAMETER ENTRY
CONFIGURATION
LANGUAGE
ERROR DISPLAY
SERVICE
DATA MEMORY 1
DATA MEMORY 2