Wiring
18
4.3
Switching contacts
One-channel version
Two-channel version
!
Note!
Condition fulfilled means:
• AV 1: concentration > Alarm value 1
• AV 2: concentration > Alarm value 2
• Fault: error occurred
Contacts AV 1, AV 2 and fault are only affected during automatic operation.
4.4
Serial interface
Connection
Terminal connection for
condition fulfilled
Terminal connection for
condition not fulfilled
Terminal connection for
power off
AV 1
A:
R:
25
25
-
-
23
24
A:
R:
25
25
-
-
24
23
25
-
24
AV 2
A:
R:
28
28
-
-
26
27
A:
R:
28
28
-
-
27
26
28
-
27
Fault
A:
R:
19
19
-
-
17
18
A:
R:
19
19
-
-
18
17
19
-
18
Unassigned
22
16
13
-
-
-
20
14
11
22
16
13
-
-
-
21
15
12
22
16
13
-
-
-
21
15
12
Connection
Terminal connection for
condition fulfilled
Terminal connection for
condition not fulfilled
Terminal connection for
power off
AV 1 - 1
A:
R:
25
25
-
-
23
24
A:
R:
25
25
-
-
24
23
25
-
24
AV 1 - 2
A:
R:
13
13
-
-
11
12
A:
R:
13
13
-
-
12
11
13
-
12
AV 2 - 1
A:
R:
28
28
-
-
26
27
A:
R:
28
28
-
-
27
26
28
-
27
AV 2 - 2
A:
R:
16
16
-
-
14
15
A:
R:
16
16
-
-
15
14
16
-
15
Fault
A:
R:
19
19
-
-
17
18
A:
R:
19
19
-
-
18
17
19
-
18
Channel ½
measurement end
A:
R:
22
22
-
-
20
21
A:
R:
22
22
-
-
21
20
22
-
21
A = NO current configured
R = NC current configured
RS 232 of CA 71
COM 1/2 at PC
SUB-D, nine-pin
Function
Function
SUB-D, nine-pin
3
TxD
RxD
2
2
RxD
TxD
3
8
CTS
RTS
7
CTS
8
5
GND
GND
5
Software protocol:
9600, N, 8, 1
Output format:
ASCII
Содержание Stamolys CA71NO
Страница 2: ......
Страница 54: ...54...