Wiring
17
4.2.2
One-channel version
4.2.3
Two-channel version
Connection
Designation
Function
Signal inputs
Leak
Liquid has collected in the drip pan
No sample
No sample available, measurement is not started, display
flashes
Signal outputs
AV 1
Alarm value 1 exceeded or undershot
AV 2
Alarm value 2 exceeded or undershot
Fault
Retrieves error message using operation menu
Measurement end
Displays "Measurement finished" (5 s)
Analog output
l-1 channel 1
0 or 4 mA = measuring range start
20 mA = measuring range end
Connection
Designation
Function
Signal inputs
Leak
Liquid has collected in the drip pan
No sample
No sample available, measurement is not started,
display flashes
Signal outputs
AV 1-1
Alarm value 1, channel 1 exceeded or undershot
AV 2-1
Alarm value 2, channel 1 exceeded or undershot
AV 1-2
Alarm value 1, channel 2 exceeded or undershot
AV 2-2
Alarm value 2, channel 2 exceeded or undershot
Fault
Retrieves error message using operation menu
Channel ½ or measurement
end
1
1) Alternative
selection
Displays active channel
Displays "Measurement finished" (5 s)
Analog output
l-1 channel 1
0 or 4 mA = measuring range start
20 mA = measuring range end
l-2 channel 2
0 or 4 mA = measuring range start
20 mA = measuring range end
Channel selection
Meas. 2
0 V = channel 1
24 V = channel 2