Installation
StamoLys CA 71 MN
10
Hauser
One-channel version
Two-channel version
•
In scope of delivery only a collecting vessel without level measurement
•
Order a second collecting vessel seperately
•
Only one collecting vessel can be mounted at housing
•
Level measurement not available
Adjusting the level measurement (one-channel version only)
Adjust the conductive level measurement due to the number of connected analyzers.
1.
In dependence of the application, mount the right adjusting pin or no adjusting pin
(Fig. 6 and Fig. 7, position 2).
2.
Pull the marked pipe (position 3) downwards due to your application (1, 2 or 3
analyzers).
E+H Collecting vessel
(at CA 71, with or without level measurement)
Connection
hose ID 3.2 mm (0.13")
Customer collecting vessel
Connection
hose ID 1.6 mm (0.06")
Max. distance from collecting vessel to analyzer
1 m (3.28 ft)
Max. height difference from collecting vessel to
analyzer
0.5 m (1.64 ft)
C07-CA71xxx-11-08-00-xx-004.eps
Fig. 6:
One Analyzer
1
M 3x12 (0.47")
2
M 3x35 (1.38")
3
Mark 1
C07-CA71xxx-11-08-00-xx-003.eps
Fig. 7:
Two analyzers
1
M 3x12 (0.47")
2
M 3x20 (0.79")
3
Mark 2
C07-CA71XXx-11-08-02-xx-001.eps
Fig. 8:
Three analyzers
1
M 3x12 (0.47")
3
Mark 3
1
2
3
1
2
3
1
3