StamoLys CA 71 MN
Operation
Hauser
23
5.3.4
PARAMETER ENTRY
!
Note!
In the following table and in the tables in the next chapter,
example
images can be
found for each option under "Display". In addition to the numerical values, the parameter
is also displayed in some options. This is
not
shown in the images. Moreover, individual
numerical values in the images can differ from the actual settings.
The
actual factory settings
can in any case be found in column 2 "Range of
adjustment / Factory settings" in bold.
Option
Range of adjustment
(default settings in bold)
Display
Info
Measuring range
Start 1
MN-A: 1 ... 150 µg/l /
0 µg/l
MN-B: 10 ... 2000 µg/l /
0 µg/l
The specified concentration is allocated a value of
0 or 4 mA at analog output 1.
Measuring range
Start 2
Two-channel version only!
The specified concentration is allocated a value of
0 or 4 mA at analog output 2.
Measuring range
End 1
MN-A: 1 ... 150 µg/l /
150 µg/l
MN-B: 10 ... 2000 µg/l /
2000 µg/l
The specified concentration is allocated a value of
20 mA at analog output 1.
Measuring range
End 2
Two-channel version only!
The specified concentration is allocated a value of
20 mA at analog output 2.
Alarm value
AV 1 - 1
MN-A: 1 ... 150 µg/l /
75 µg/l
MN-B: 10 ... 2000 µg/l /
1000 µg/l
Concentration threshold value limit relay 1,
channel 1 (differential hysteresis 2% of alarm
value).
Alarm value
AV 2 - 1
MN-A: 1 ... 150 µg/l /
150 µg/l
MN-B: 10 ... 2000 µg/l /
2000 µg/l
Concentration threshold value limit relay 2,
channel 1 (differential hysteresis 2% of alarm
value).
Alarm value
AV 1 - 2
MN-A: 1 ... 150 µg/l /
75 µg/l
MN-B: 10 ... 2000 µg/l /
1000 µg/l
Two-channel version only!
Concentration threshold value limit relay 1,
channel 2 (differential hysteresis 2% of alarm
value).
Alarm value
AV 2 - 2
MN-A: 1 ... 150 µg/l /
150 µg/l
MN-B: 10 ... 2000 µg/l /
2000 µg/l
Two-channel version only!
Concentration threshold value limit relay 2,
channel 2 (differential hysteresis 2% of alarm
value).
Time
1st measurement
01.01.96 00:00... 31.12.95 23:59
Date format DD.MM.YY, time hh.mm. After each
change the instrument does not wait for the
measuring interval. If the measurement is to start
immediately, set the time in the past.
Measuring interval
6 ... 120 min
10
Time between two measurements.
If the setting is 2 minutes, the measurements take
place without any pauses.
Frequency of
measurement
Channel 1
0 ... 9
1
1
Two-channel version only!
Number of measurements at channel 1 before
switching to channel 2.
Frequency of
measurement
Channel 2
0 ... 9
1
1
Two-channel version only!
Number of measurements at channel 2 before
switching to channel 1.
Range start 1
0.00 mg/l
Range start 2
0.00 mg/l
Range end 1
2.50 mg/l
Range end 2
2.50 mg/l
Alarm val. 1-1
2.50 mg/l
Alarm val. 2-1
1.25 mg/l
Alarm val. 1-2
1.25 mg/l
Alarm val. 2-2
2.50 mg/l
1. Measurement
10.02.02 08:00
Meas. interval
10 min
n* Channel 1:
9
n* Channel 2:
1