StamoLys CA 71 MN
Operation
Hauser
25
5.3.5
LANGUAGE
The following languages are available:
•
Deutsch
•
English
•
Français
•
Nederlands
•
Suomi
•
Magyar
•
Polski
•
Italiano.
5.3.6
ERROR DISPLAY
!
Note!
•
This menu is a "Read-Only-Menu".
•
You can find the individual error messages, their meaning and solutions to problems
in chapter "Trouble-shooting instructions".
•
If there is at least one error message, the signal output is set to "fault".
•
Causes of faults are requested for every measurement. If an error which occurred
previously no longer exists, it is automatically cancelled. If this should not happen
automatically, error messages can be deleted by quickly switching the analyzer off
and back on again.
5.3.7
SERVICE
Navigate through the main menu with the
W
key to the "SERVICE" option and press the
F
key to enter the menu.
Option
Display
Info
Pumps and valves
"Virtual switching board"
Various valve and pump combinations can be
selected.
The setting options are:
– Valve 1:
P (sample) or S (standard)
– Pump 1 and pump 2:
s (stop) or g (go)
– Valve 2:
S (standard) or C (cleaning agent)
– Valve 3 (
Two-channel version only
):
1 (channel 1) or 2 (channel 2)
The following valve combinations are possible:
(applies to one-channel and two-channel version,
whereby where the latter is concerned selection is
made by positioning valve 3 between channel 1
and 2)
– V1: P, V2: S
Passage for the sample. This combination is
automatically reset on leaving the service menu.
– V1: S, V2: S
Passage for standard solution
– V1: S, V2: R
Passage for cleaning agent
Signal frequency
Signal frequency of the photometer
Mixture
The reagent and sample pumps can be switched
on together, so that they run in the same ratio as for
filling the sample-reagent mixture in measurement
mode. Selection between s (stop) and g (go).
V1 P1 P2 V2 V3
P
s
s
S
1
0 Hz
Mix
s