Prosonic T - FMU 130, 131, 230, 231, 232
Hauser
yes
yes
Expected
Actual
Expected
Actual
➃
Measured value jumps
Is the signal affected
– Increase output damping
sporadically with
by turbulence or
see page 24.
constant level and
agitator blades?
– With agitator blades in measuring
turbulence or agitator
range, check sensor position.
blades.
see page 8-11, 25.
– Select application parameter
0 or 2 in V0H3.
See page 21.
➄
The measured value
Multiple echoes?
– Select application parameter 2
jumps to a lower value
in V0H3.
or remains continuously
See page 21.
too low with constant
– Select a larger first echo factor 1
level.
or 2 in V3H4.
See page 26.