Prosonic T - FMU 130, 131, 230, 231, 232
Simulation
The simulation mode enables Prosonic T functions to be simulated and checked.
Errors and Warnings in V9H0
•
E613
: Display during simulation.
Returns to normal operation after simulation.
Simulation Off: V9H6: 0
• On power failure the instrument automatically returns to normal operation!
Simulation of Height
Locking
Locking via the Keyboard
When the instrument is locked via the keyboard, both keyboard and display programming
as well as all programming via the handheld terminals FMX 770, FXN 671 etc. are blocked.
It can only be unlocked using the keyboard.
The matrix can again be locked once all parameters have been entered.
• Locking by entering a three-figure code number not equal to 333.
• Locking by using the keyboard (see note on locking via the keyboard).
#
VH
Entry
Text
1 V9H6
1
H
Simulation
height
2 V9H7
e.g. 6.600 ft.
H
Simulated
(2.000)
height
(e.g. 6.6 ft./2 m)
3 V9H8
Display of current (also shown on
bargraph)
V0H0
Display of height, level or volume
4 V9H6
0
H
Simulation off
#
VH
Entry
Text
1 V9H6
3
H Simulation
current
2 V9H7
e.g. 14
H Simulated
current
(e.g. 14 mA)
3 V9H8 Display of current (also
shown on bargraph)
4 V9H6 0
H Simulation off
Simulation of Current
#
VH
Entry
Text
1 V9H6
2
H Simulation volume
2 V9H7
e.g. 26.40 gal.
H Simulated
(100.0)
volume
(e.g. 26.4 gal/100 l)
3 V9H8 Display of current (also shown on
bargraph)
V0H0 Display of volume (if no linearization
curve has been entered, the volume
corresponds to level).
4 V9H6
0
H Simulation off
Simulation of Volume
#
VH
Entry
Text
1
V9H9
333
H
Unlocking
2
The number 333 is shown in V9H9.
Matrix fields are no longer blocked.
#
VH
Entry
Text
1
V9H9
e.g. 332
H
Locking
2
The number 332 is shown in V9H9.
All matrix fields are blocked except V9H9.
P = Protect
F = Free
Unlocking
display approx.
2 s
333 displayed
in V9H9
Locking
display approx.
2 s
9999 displayed
in V9H9