Prosonic T - FMU 130, 131, 230, 231, 232
Hauser
Operating via the Matrix
The Prosonic T is calibrated and operated using the 10 x 10 Hauser user
matrix.
The
basic configuration
can be realized in its simplest form over
three matrix fields
.
Configuration and operation are always identical for:
• Local push-button operation with display
• Operation via handheld terminal
• Operation via the Silometer FMX 770 (BA 136F) or FXN 671 (TI 236F)
• Operation via a process bus
Local Push-button Operation with Plug-In Display
Display Elements
VH position
Parameter
Bar display:
– Display: current or echo quality
– flashes on alarm
Push Buttons
Push-button
Function
Selecting the Matrix Field
Selecting the vertical matrix position.
Selecting the horizontal matrix position.
and
The display jumps to V0H0.
Entering Parameters
or
Activates the appropriate matrix position. The selected
position flashes.
Changes the value of the flashing position by +1.
Changes the value of the flashing position by –1.
and
Resets the value entered to the original value if it is not yet
confirmed with
V
or
H
.
Confirming the Entry
or
Confirms the entry by leaving the matrix field.
Locking/Unlocking the Matrix
and
Locking matrix, 9999 is shown in V9H9.
and
Unlocking matrix, 333 is shown in V9H9.