Prosonic Flow 92
11 Description of device functions
Hauser
69
Site Name
An empty line in column “SITE NAME” indicates that there is no data set
available at the respective number. Start defining a new measurement site in
the function group “SITE DATA” (see Page 69).
Site Setup
Site Data
The function group “SITE DATA” serves for entry of all site data and assists during commissioning of the
measurement.
Site Name (Entry)
Choose this function for entering the name of the measurement site.
An alphanumeric on-screen keyboard will assist you during entry. Choose
“EXIT” to complete the entry and close the on-screen keyboard.
Pipe Circumference
Use this function to enter the outer circumfernce of the pipe.
User input (Metric):
Fixed-point number 40.00...19200.00 mm
User input (US):
Fixed-point number 1.5000...760.0000 inch
Factory setting:
190.06 mm / 7.5000 inch
Pipe Diameter
Use this function to enter the outer diameter of the pipe.
User input (Metric):
Fixed-point number 13.00...6100.00 mm
User input (US):
Fixed-point number 0.5000...240.0000 inch
Factory setting:
60.0 mm
Pipe Material
Use this function to enter the pipe material.
Options:
CARBON STEEL – STAINLESS STEEL – CAST IRON – COPPER – PVC –
ALUMINIUM – DUCTILE IRON – ASBESTOS CEMENT – GRP – PEEK – PVDF
– ACRYLIC GLASS – OTHER
Note!
• If “OTHER” is selected, the sound velocity can be entered in m/s (metric)
or ft/s (US).
• If a liquid apart from “OTHER” is selected, the temperature of the fluid
must be entered.
Factory setting:
PVC
Site Setup
Parameter Memory
The function group “PARAMETER MEMORY” serves for entry and recovery of all site data and assists
during commissioning of the measurement.
Содержание prosonic flow 92
Страница 4: ...Brief operating instructions Prosonic Flow 92 4 Endress Hauser...
Страница 12: ...2 Identification Prosonic Flow 92 12 Endress Hauser...
Страница 26: ...3 Installation Prosonic Flow 92 26 Endress Hauser...
Страница 34: ...4 Wiring Prosonic Flow 92 34 Endress Hauser...
Страница 40: ...5 Operation Prosonic Flow 92 40 Endress Hauser...
Страница 46: ...7 Maintenance Prosonic Flow 92 46 Endress Hauser...
Страница 96: ...Index Prosonic Flow 92 96 Endress Hauser...