Prosonic Flow 92
10 Technical data
Hauser
57
Maximum measured error
For flow velocities between 0.5 m/s and 7 m/s and a Reynolds number of >10000, the
system accuracy is:
± 0.5 % o.r. (of current reading)
The system is dry calibrated. The calibration factor by dry calibration is calculated
based on the actual pipe and liquid properties. This dry calibration procedure results in
an additional uncertainty for the measurement. The resulting accuracy of the measure-
ment therefore is better than 2 % typically.
The zero point instability is < 10 mm/s.
Repeatability
Max. ± 0.3% for flow velocities > 0.5 m/s
10.1.7
Operating conditions
Installation
Installation instructions
Any orientation (vertical, horizontal)
Restrictions and additional installation instructions
Inlet and outlet run
Length of connecting
cable
Shielded cables are offered in the following lengths:
5 m and 10 m
!
Note!
Route the cable well clear of electrical machines and switching elements.
Environment
Ambient temperature
range
• Transmitter Prosonic Flow 92:
−
10...+45°C
• Flow measuring sensors Prosonic Flow W, U:
–20...+60 °C
• Flow measuring sensors Prosonic Flow P:
0...+170 °C
• Sensor cable PVC:
–20...+70 °C
• Sensor cable PTFE (only for Prosonic Flow W/P):
–40...+170 °C
Avoid direct sunlight, particularly in warm climatic regions.
Storage temperature
The storage temperature corresponds to the operating temperature range of the trans-
mitter and the appropriate flow measuring sensors and the corresponding sensor cable
(see above).
Содержание prosonic flow 92
Страница 4: ...Brief operating instructions Prosonic Flow 92 4 Endress Hauser...
Страница 12: ...2 Identification Prosonic Flow 92 12 Endress Hauser...
Страница 26: ...3 Installation Prosonic Flow 92 26 Endress Hauser...
Страница 34: ...4 Wiring Prosonic Flow 92 34 Endress Hauser...
Страница 40: ...5 Operation Prosonic Flow 92 40 Endress Hauser...
Страница 46: ...7 Maintenance Prosonic Flow 92 46 Endress Hauser...
Страница 96: ...Index Prosonic Flow 92 96 Endress Hauser...