The measuring equipment comprises:
•
Transmitter Promag 31 and
•
Sensor Promag F
Remote-mounted version (FS version or FL version)
The transmitter is mounted remote from the sensor:
FS version
•
Up to 10 m distance
meduim conductivity
min. 5
µ
S/cm
•
From 10
…
200 m distance
max. cable length in terms
of the medium conductivity
(5...200
µ
S/cm)
FL version
•
Max. cable length 200 m, not dependent on
conductivity.
•
No EPD available.
•
The electrical connection between transmitter and
sensor is made in the connection housing
Compact version:
The Promag 31 transmitter
and the sensor together form
a mechanical unit.
Promag F
Promag 31
Promag 31 F
b
a028
y
0
3
Fig. 3
Promag 31
1. System Description
Hauser
7