2.2 Transport instructions for Promag from DN 350 / 14"
The pipe lining on the flanges is protected by disks to prevent damage when
transporting to the installation point. These are to be removed before installation.
The instruments are to be transported in the containers they are delivered in.
Transporting to the measuring point
The grips on the flange must be used for
lifting out and mounting the sensor in the
piping.
The sensor should not be lifted by the
connection housing.
The sensor should not be lifted by the
sheet casing using a forklift truck.
This can buckle the casing and damage
the internal magnetic coils.
Base of the sensor
The sensor should stand on a base
which is sufficiently strong to withstand
its weight.
Note!
The sensor must not be supported by
the sheet casing! This can buckle the
casing and damage the internal mag-
netic coils.
b
a02
8y
6
0
Fig. 6
b
a
028y
61
Fig. 7
b
a028
y
6
2
Fig. 8
b
a028
y
6
4
Fig. 9
b
a028
y
6
3
2. Mounting and Installation
Promag 31
12
Hauser