Display functions
Display
Function
Description
tot
Display totaliser
Display of the current flow rate or totaliser volume.
A negative flow direction is indicated by a negative
digit.
To select the measuring unit
⇒
press “
Set
” key
rate
Display flow rate
only in
non fiscal
operation!
Caution!
Setting of SI/US units is done by miniature switches on
the measuring amplifier board (see Chapter 4.2)
.
dISP-OF
Display
overflow totalis er
(is only possible if
you break the seal)
Display of the number of overflows at numerical values
>
99 999 999
.
Note!
A maximum of 21 overflows is displayed. Starting with
the 22nd overflow, the totaliser begins to sum up from
“0” upwards.
rAtE-tot
Display
Flow rate/totaliser
Only with
non fiscal
measurement operation!
Alternating display (about every 10 seconds) of the
current flow rate and totaliser value.
EPd-oFF
Empty Pipe
Detection (EPD)
The EPD function detects whether a measuring pipe is
only partially filled with liquid
For ON/OFF switching
⇒
press “
Set
” key
EPd-Ad_E
Empty pipe
adjustment
Only with
non fiscal
measurement operation!
Empty/full pipe adjustment for EPD
To start adjustment
⇒
press “
Set
” key
EPd-Ad_F
Full pipe adjustment
Note!
•
Any adjustment has to be done
before
switching
on EPD (otherwise the
ADJ_ERROR
message is
displayed).
•
During adjustment the message
ADJ_BUSY
is
displayed for about 0.5 s. After any adjustment
ADJ_DONE
is displayed
tESt
Test function
This function can only be activated with fiscal
measurement status over the auxiliary input!
Accessing this function activates an automatic test
sequence of all display elements; the following displays
are shown:
1.
+ 88 888 888
(incl. display segments)
2.
- 00 000 000
(without display segments)
3. All display elements are blank
4. Flow indication
Additional Information!
•
In case of system or process errors (incl. EPD) outputs react as described in Chapter 5.1 of the
Operating Manual.
•
In case of a supply breakdown, all measuring data (e.g. totaliser value) and configurations are
saved and again available once the device is restarted.
•
If a sensor equipped with an EPD electrode has to be exchanged during servicing,
then the EPD calibration must always be carried out again.
Promag 31
4. Operation and Commissioning
Hauser
37
Содержание promag 31
Страница 4: ...Promag 31 4 Endress Hauser ...
Страница 22: ...2 Mounting and Installation Promag 31 22 Endress Hauser ...
Страница 28: ...Promag 31 28 Endress Hauser ...
Страница 44: ...5 Fault Location and Remedies Promag 31 44 Endress Hauser ...
Страница 56: ...6 Technical Data Promag 31 56 Endress Hauser ...
Страница 59: ......