Proline Promass 80E, 83E
42
Hauser
SMS 1145 (threaded hygienic connection)
A0021291
Dimension unit in mm (in)
Rupture disc
Sensor housings with integrated rupture disks are optionally available.
#
Warning!
• Make sure that the function and operation of the rupture disk is not impeded through the
installation. Triggering overpressure in the housing as stated on the indication label. Take adequate
precautions to ensure that no damage occurs, and risk to human life is ruled out, if the rupture disk
is triggered.
Rupture disk: Burst pressure 10 to 15 bar (145 to 217.5 psi).
• It is not permitted to open the connections or remove the rupture disk.
"
Caution!
Rupture disks can not be combined with separately available heating jacket.
!
Note!
• Please note the indication labels.
A0007823
Indication label for the rupture disk
The position of the rupture disk is indicated by an adhesive label on top of the disk. If the rupture disk
is triggered, the adhesive label is damaged and can thus be visually monitored.
di
G
U
E
+1.5
–2.0
L
(+0.06)
(–0.08)
Threaded hygienic connection SMS 1145:
1.4404/316L
DN
G
L
U
di
3/8"
Rd 40 × 1/6"
9.02
0.89
0.21
½"
Rd 40 × 1/6"
10.75
0.89
0.33
1"
Rd 40 × 1/6"
12.76
0.89
0.47
1½"
Rd 60 × 1/6"
17.95
1.40
0.69
2"
Rd 70 × 1/6"
22.13
1.91
1.02
3"
Rd 98 × 1/6"
26.46
2.87
1.59
3A version also available (Ra
30
μ
in/150 grit.); All dimensions in [inch]
RUPTUR
E
DISK
i
Содержание Proline Promass 80E
Страница 49: ...Proline Promass 80E 83E Endress Hauser 49...