Proline Promass 80E, 83E
22
Hauser
Flange connection to DIN 11864-2 Form A (flat flange with groove)
Flange material: 1.4404/316L
A0020978-EN
Threaded hygienic connection to ISO 2853
Connection material: 1.4404/316L
A0020988-EN
Rupture disk
The sensor housing protects the inner electronics and mechanics and is filled with dry nitrogen. The
housing of this sensor does not fulfill any additional secondary containment function. However, 15 bar
(217.5 psi) can be specified as a reference value for the pressure loading capacity.
For increased safety, a version with rupture disk (triggering pressure 10 to 15 bar / 145 to 217.5 psi)
can be used, which is available for order as a separate option.
Further information
ä
42.
Limiting flow
See information in the "Measuring range" section
ä
5.
Select nominal diameter by optimizing between required flow range and permissible pressure loss. See
the "Measuring range" section for a list of maximum possible full scale values.
• The minimum recommended full scale value is approx. 1/20 of the max. full scale value
• In most applications, 20 to 50% of the maximum full scale value can be considered ideal
• Select a lower full scale value for abrasive substances such as fluids with entrained solids
(flow velocity <1 m/s (<3 ft/s))
• For gas measurement the following rules apply:
– Flow velocity in the measuring tubes should not be more than half the sonic velocity (0.5 Mach)
– The maximum mass flow depends on the density of the gas: formula
ä
5
Pressure loss
To calculate the pressure loss, use the
Applicator
sizing tool (
ä
47).
DN 8...40
DN 50…80
300
[°F]
350
250
200
150
100
50
0
-50
[°C]
200
150
100
50
0
-50
400
0
10
20
30
40
200
400
0
[bar]
[psi]
300
[°F]
350
250
200
150
100
50
0
-50
[°C]
200
150
100
50
0
-50
400
0
10
20
30
[bar]
[psi]
200
400
0
Содержание Proline Promass 80E
Страница 49: ...Proline Promass 80E 83E Endress Hauser 49...
Страница 50: ...www addresses endress com...