3. Pull out the display module with a gentle rotational movement. To make it easier to access
the lock switch, attach the display module to the edge of the electronics compartment.
Ã
4. Setting the write protection switch (WP) on the main electronics module to the ON position
enables the hardware write protection.
Ã
The
Hardware locked
option is displayed in the
Locking status
parameter. In
addition, on the local display the -symbol appears in front of the parameters in the
header of the operational display and in the navigation view.
5. Feed the spiral cable into the gap between the housing and main electronics module and
plug the display module into the electronics compartment in the desired direction until it
engages.
6. Reverse the removal procedure to reassemble the transmitter.
11
Diagnostic information
Faults detected by the self-monitoring system of the measuring device are displayed as a
diagnostic message in alternation with the operational display. The message on remedial
measures can be called up from the diagnostic messages, and contains important information on
the fault.
4
2
5
X X X X
X X X
X
X
X X X X
X X X
X
X
S441
S
S
mA
i
1
3
20.50
S441
2d11h05m55s
(ID:153)
1. Check process
2. Check current output settings
Curr. output 1
Curr. output 1
A0013940-EN
å
11
Message for remedial measures
1
Short text
2
Diagnostic behavior with diagnostic code
3
Service ID
4
Operation time of occurrence
5
Remedial measures
The user is in the diagnostic message.
1. Press
O
( symbol).