Proline Promag H 200
Installation
Hauser
19
6.2.2
Preparing the measuring device
1. Remove all remaining transport packaging.
2. Remove any protective covers or protective caps present from the sensor.
3. Remove stick-on label on the electronics compartment cover.
6.2.3
Mounting the sensor
!
WARNING
Danger due to improper process sealing!
►
Ensure that the inside diameters of the gaskets are greater than or equal to that of the process
connections and piping.
►
Ensure that the gaskets are clean and undamaged.
►
Install the gaskets correctly.
1. Ensure that the direction of the arrow on the sensor matches the flow direction of the
medium.
2. To ensure compliance with device specifications, install the measuring device between the
pipe flanges in a way that it is centered in the measurement section.
3. Install the measuring device or turn the transmitter housing so that the cable entries do
not point upwards.
Ã
A0013964
The sensor is supplied to order, with or without pre-installed process connections. Pre-installed
process connections are secured to the sensor with 4 or 6 hex-head threaded fasteners.
The sensor might require support or additional attachments, depending on the application
and the length of the piping run. When plastic process connections are used, the sensor
must be additionally supported mechanically. A wall-mounting kit can be ordered
separately from Hauser as an accessory .