3 Transducer Block
Device functions Proline Promag 55 FOUNDATION Fieldbus
94
Hauser
3.5
Parameters of the "Display" Transducer Block
The following table shows all the Hauser-specific parameters of the "Display" Transducer
Block. These can only be changed after entering an enabling code in the "Access - Code"
parameter.
!
Note!
FOUNDATION Fieldbus parameters are described in the "FOUNDATION Fieldbus Overview"
Operating Instructions (BA013S) (available at:
→
www.endress.com
→
Download).
"Display" Transducer Block/base index 1800:
Parameter
Write access with
operating mode
(MODE_BLK)
Description
Access - Code
AUTO - OOS
All data of the measuring system are protected against inadvertent change.
Only when the code has been entered in this parameter can the
manufacturer-specific parameters be programmed and the device
configuration modified.
You can enable programming by entering:
• Code 55 (factory setting)
• Private code (
User input:
Max. 4-digit number (0 to 9999)
!
Note!
• If the write protection is enabled then access to the manufacturer-
specific parameters is blocked even if the right code is entered. Write
protection can be activated and deactivated by means of jumpers on the
I/O board (
→
see Operating Instructions for Proline Promag 55
FOUNDATION Fieldbus).
• You can disable programming again by entering any number (other than
the access code) in this parameter.
• The Hauser service organization can be of assistance if you
lose your private code.
• Certain parameters are not accessible unless a special service code is
entered. This service code is known only to the Hauser service
organization. Please contact your Hauser service center if you
require clarification.
• The entry made here does not affect the local display. Thus,
programming via the function matrix must be enabled separately.
Access - Status
read only
Use this parameter to display the current status of the possibilities for
accessing the manufacturer-specific parameters of the device.
User interface:
• LOCKED (parameterization disabled)
• ACCESS CUSTOMER (parameterization enabled)
• ACCESS SERVICE (parameterization enabled, access to service level)
Access - Code
Counter
read only
Displays how often the customer code, service code or the digit "0"
(code free) has been entered to gain access to the function matrix.
User interface:
Max. 7-digit number: 0 to 9999999
Factory setting:
0
Содержание Proline Promag 55
Страница 60: ...8 Block SUPERVISION Device Functions Proline Promag 55 FOUNDATION Fieldbus 60 Endress Hauser ...
Страница 66: ...10 Index local operation Device Functions Proline Promag 55 FOUNDATION Fieldbus 66 Endress Hauser ...
Страница 68: ...Table of contens FOUNDATION Fieldbus Device Functions Proline Promag 55 FOUNDATION Fieldbus 68 Endress Hauser ...
Страница 129: ...Device Functions Proline Promag 55 FOUNDATION Fieldbus Endress Hauser 129 ...
Страница 130: ...www endress com worldwide BA127D 06 en 09 06 71032705 FM SGML 6 0 ...