Device Functions Proline Promag 55 FOUNDATION Fieldbus
7 Block BASIC FUNCTION
Hauser
45
PRESSURE SHOCK
SUPPRESSION
(6404)
The closure of a valve can cause brief but severe movements of the fluid in the piping
system, movements which the measuring system registers. The pulses totaled in this way
result in a totalizer reading error, particularly in the case of batching processes. For this
reason, the measuring device is equipped with pressure shock suppression (= short-term
signal suppression) which can eliminate system-related "disruptions".
!
Note!
Note that pressure shock suppression cannot be used unless the low flow cut off is active
(see ON-VALUE LOW FLOW CUT OFF function on Page 44).
Use this function to define the time span for active pressure shock suppression.
Activation of pressure shock suppression
Pressure shock suppression is activated once the flow falls below the switch-on point of
the low flow cut off (see point
a
in graphic).
While pressure shock suppression is active, the following conditions apply:
• Flow reading on display
→
0
• Totalizer reading
→
The totalizers are pegged at the last correct value.
Deactivation of pressure shock suppression
The pressure shock suppression is deactivated after the time interval, set in this function,
has passed (see point
b
).
!
Note!
The actual flow value is displayed and output, when the time interval for the pressure
shock suppression has passed and the flow exceeds the switch-off point of the low flow
cut off (see point
c
in graphic).
A0001285-EN
m
= on value (low flow cut off),
n
= off value (low flow cut off)
a
Activated when the on-value for low flow cut off is undershot
b
Deactivated once the time specified expires
c
Flow values are taken into account again for calculating the pulses
■
Suppressed values
Q
Flow
User input:
Max. 4-digit number, incl. unit: 0.00 to 100.0 s
Factory setting:
0.00 s
Function description
BASIC FUNCTION
→
PROCESS PARAMETER
→
CONFIGURATION
n
m
a
b
c
inactive
active
inactive
t
Pressure shock suppression
Command:
Valve closes
Q
Given time
After filling
Содержание Proline Promag 55
Страница 60: ...8 Block SUPERVISION Device Functions Proline Promag 55 FOUNDATION Fieldbus 60 Endress Hauser ...
Страница 66: ...10 Index local operation Device Functions Proline Promag 55 FOUNDATION Fieldbus 66 Endress Hauser ...
Страница 68: ...Table of contens FOUNDATION Fieldbus Device Functions Proline Promag 55 FOUNDATION Fieldbus 68 Endress Hauser ...
Страница 129: ...Device Functions Proline Promag 55 FOUNDATION Fieldbus Endress Hauser 129 ...
Страница 130: ...www endress com worldwide BA127D 06 en 09 06 71032705 FM SGML 6 0 ...