Micropilot M FMR 240 with Foundation Fieldbus
Mounting
E Hauser
25
Example for the construction of a bypass.
L00-FMR2xxxx-17-00-00-en-019
100 %
0 %
Micropilot M
FMR230,
FMR240,
FMR245
full bore
ball valve
e.g. welding neck flange
DIN 2633
Recommendation:
min.
400 mm / 160"
Diameter of o
of ball valve must always
be equivalent to pipe diameter.
Avoid edges and constrictions.
pening
Do not weld
through the pipe wall.
The inside of
the bypass
must remain
smoth.
Diameter of
the connection pipes
as small as possible
20 cm
8”
20 cm
8”
20 cm
8”
20 cm
8”
gap
1 mm
0.04”
£ £
marker