Micropilot FMR51, FMR52 HART
Diagnostics and troubleshooting
Hauser
75
Error
Possible cause
Remdy
If measured distance (Setup
→ Distance) does not match the real
distance:
Interference echo
Carry out tank mapping (
Confirm
distance
parameter (→ 111)).
No change of measured value
on filling / emptying
Interference echo from installations,
nozzle or build-up on the antenna.
• Carry out tank mapping (
Confirm
distance
• If necessary, clean antenna
• If necessary, selet better mopunting
position
If the surface is not calm (e.g.
filling, emptying, agitator
running), the measured value
jumps sporadically to a
higher level
Signal is weakened by the rough
surface - the interference echoes are
sometimes stronger.
• Carry out tank mapping (
Confirm
distance
• Select
=
Process vessel with agitator
.
• Increase integration time (Expert
→ Sensor → Distance → Integration
time)
• Optimize orientation of the antenna
• If necessary, select a better mounting
position and/or larger antenna.
During filling/emptying the
measxured value jumps
downwards
Multiple echoes
• Check
Tank type
parameter
• If possible, do not select central
installation position.
• If appropriate, use a stilling well.
Error message F941 or S941
"Echo lost"
Level echo is too weak.
• Check
Medium group
parameter
• If necessary, select a more detailed
setting in
Medium property
• Optimize alignment of antenna
• If necessary, select a better
installation position and/or larger
antenna.
Level echo suppressed.
Delete mapping and record it again.
Device displays a level when
the tank is empty.
Interference echo
Carry out mapping over entire
measuring range when the tank is
empty (
Confirm distance
parameter
Wrong slope of the level in
the entire measuring range
Wrong tank type selected.
Set
Tank type
correctly.