Micropilot FMR51, FMR52 HART
Operating menu
Hauser
177
16.4.7 "Device check" submenu
Navigation
Diagnostics → Device check
Start device check
Navigation
Diagnostics → Device check → Start dev. check
Description
Start a device check.
Selection
• No
• Yes
Factory setting
No
Additional information
In the case of a lost echo a device check can not be performed.
Result device check
Navigation
Diagnostics → Device check → Result dev.check
Description
Displays the result of the device check.
Additional information
Meaning of the display options
•
Installation ok
Measurement possible without restrictions.
•
Accuracy reduced
A measurement is possible. However, the measuring accuracy may be reduced due to the
signal amplitudes.
•
Measurement capability reduced
A measurement is currently possible. Howerver, there is the risk of an echo loss. Check
the mounting position of the device and the dielectric constant of the medium.
•
Check not done
No device check has been performed.
Last check time
Navigation
Diagnostics → Device check → Last check time
Description
Displays the operating time at which the last device check has been performed.
Содержание Micropilot FMR50
Страница 116: ...Operating menu Micropilot FMR51 FMR52 HART 116 Endress Hauser User interface Init recording In progress Finished ...
Страница 183: ......
Страница 184: ...www addresses endress com 71344017 71344017 ...