Liquisys M CUM223/253
Diagnostics and troubleshooting
Hauser
77
E162
Dosage stop
Check settings in the CURRENT INPUT
and CHECK function groups.
Yes
No
No
-
-
E171
Flow in main stream too
low or zero
Restore flow.
Yes
No
No
-
-
E172
Switch-off limit for
current input exceeded
Check process variables at sending
measuring instrument. Change range
assignment if necessary.
Yes
No
No
-
-
E173
Current input < 4 mA
Check process variables at sending
measuring instrument.
Yes
No
No
-
-
E174
Current input > 20 mA
Check process variables at sending
measuring instrument. Change range
assignment if necessary.
Yes
No
No
-
-
1)
PV = Process variable
2)
If this error occurs, there is no possibility of starting a cleaning session (field F8 not applicable with this error).
3)
Current error messages not applicable via PROFIBUS
Error
no.
Display
Tests and/or remedial measures
Alarm contact
Error current
Autom.
cleaning start
PROFIBUS
status
Facty
User
Facty
User
Facty
User