Installation
Liquisys M CUM223/253
12
Hauser
3.3
Installation instructions
3.3.1
Field instrument
There are several ways of securing the field housing:
• Wall mounting with fixing screws
• Post mounting to cylindrical pipes
• Post mounting to square securing masts
NOTICE
Effect of climate conditions (rain, snow, direct sun etc.)
Impaired operation to complete transmitter failure
‣
When installing outside, always use the weather protection cover (accessory).
Transmitter wall mounting
a0005736
Fig. 4:
Wall mounting field device
For wall mounting the transmitter, proceed as follows:
1. Drill the bores as shown in
2. Drive the two fixing screws through the securing bores (1) from the front.
3. Mount the transmitter on the wall as shown.
4. Cover the bores with plastic caps (2).
1
2
154 / 6.06
mm / inch
Ø6/ Ø