Liquisys M CUM223/253
Commissioning
Hauser
67
Version TB: Initial settings for residual concrete water
The version TB contains the Plus-package with additional setting options. The default values
set in individual fields of the menus differ from the standard version to make commissioning
as simple as possible.
These values are so selected that no additional settings need to be made for applications in
residual concrete water
. If you ever reset the device to the original factory settings (set
default), you can find valid values for residual concrete water in the table below.
Calibration is carried out in the % measuring range (the transmitter switches
automatically). With the above setting, a density of e.g. 1.12 kg/l = 12 % must be
selected. In some cases, it may be necessary to adapt the calibration to a real sample. To
do so, perform a single-point calibration.
Menu field
Setting
Mode of operation
A1
spec.
Unit
A2
kg/l
Display format
A3
XX.xx
Measured value damping
A5
10
Calibration data record
B4
3
Current output
O1
Out1
Characteristic
O2
lin
Current range
O211
0 to 20 mA
Measured value 0 mA
O212
1.00
Measured value 20 mA
O213
1.30
Active characteristic
K1
1
Edited characteristic
K2
1
Table option
K3
edit
Number of value pairs
K4
2
Support point
K5
1 to 2
Measured value / display value
K6 / K7
1: 0%/1.00 2: 50%/1.50
Language version
S1
GER