Coding
Field
Selection or range
Factory setting
Display
Info
C1 (3)
Selection of
calibration
3 Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit
=
Edit
calibration (3)
Refl = Fitting
with reflection
effects (4)
1-Pt = Single-point
calibration (5)
Data =Calibration
data (6)
If the individual samples were
defined in the laboratory, you can
correct the calibration here.
C131
Enter concentra-
tion of the first
calibration
solution
Current value from
C111
entire measuring
range
C132
Enter concentra-
tion of the second
calibration
Current value from
C112
C132
≥
1.1
⋅
C131
C133
Enter concentra-
tion of the third
calibration
solution
Current value from
C113
C133
≥
1.1
⋅
C132
C134
Calibration status
is displayed
o.k.
E xxx
C135
Store calibration
results
yes
no
new
If C134 = E xxx, then only no or
new
(Exception: calibration
warning E 84).
If new, return to C.
If yes/no, return to
“Measurement”.
C1 (4)
Selection of
calibration
3 Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit
calibration (3)
Refl
=
Fitting
with reflection
effects
(4)
1 Pt = Single-point
calibration (5)
Data =Calibration
data (6)
Only for solutions
≤
2 FNU /
5 ppm!
There is no effect of the
installation in clear media
Factory settings are printed in
bold
face;
base version does not include functions in
italic.
Edit
C1
Calibrat
100.0
C131
Concentr1
FNU
330.0
C132
Concentr2
FNU
1000.0
C133
Concentr3
FNU
o.k.
C134
Status
yes
C135
Store
Refl
C1
Calibrat
UM253E5B.CHP
Liquisys M CUM 223 / 253
Instrument configuration
Hauser
51