5.7
Service
Coding
Field
Selection or range
Factory setting
Display
Info
K5
Selection of triplet
1
1 ... Number of support
points from K4
Fin
Any triplet can be edited.
K6
Enter turbidity
value
0.0 %
Entire measuring
range
The function chain K5 ... K7 is
run through as many times as is
equal to the value in K4.
“Finished” appears as the last
step. After confirmation the
system jumps to K8.
K7
Enter
concentration
value
Entire measuring
range
Measurement unit as in A2.
K9
Enter whether or
not table status is
okay
yes
no
Return to K.
If not, then set table correctly (all
previous settings are kept) or
back to measurement mode (this
makes the table invalid).
Factory settings are printed in
bold
face;
base version does not include functions in
italic.
Coding
Field
Selection or range
Factory setting
Display
Info
S
Function group
SERVICE
S1
Selection of
language
ENG
=
English
GER = Deutsch
FRA = Francais
ITA
= Italiano
NEL = Nederlands
ESP = Espanol
This field must be set once
during device configuration. After
confirmation with ENTER
S2
Hold configuration
S+C
=
during
setup and
calibration
CAL = during
calibration
Setup = during setup
none = no hold
S = setup,
C = calibration.
K5
Sel.Elem.
1
K6
MeasValue
0
%
K7
concentr.
0
kg/l
K8
Status ok
yes
S
SERVICE
ENG
S1
Language
S+C
S2
Auto HOLD
Instrument configuration
Liquisys M CUM 223 / 253
42
Hauser