Coding
Field
Selection or range
Factory setting
Display
Info
O211
Selection of
current range
4–20 mA
0–20 mA
O212
0/4 mA value;
enter
corresponding
turbidity or
temperature
value
0.0 FNU
0.0 ppm
0.0 mg/l
0.0 g/l
0.0 kg/l
0.0 t/m3
0.0 %
0.0 °C
entire measuring
range
Enter the measured value
corresponding to the minimum
current value (0/4 mA) at the
transmitter output.
Display format from A3.
O213
20 mA value;
enter
corresponding
turbidity or
temperature
value
9999 FNU
9999 ppm / 3000 ppm
9999 mg/l / 3000 mg/l
300.0 g/l / 3 g/l
99.99 kg/l
99.99 t/m3
10.0 %
100.0 °C
entire measuring
range
Enter the measured value
corresponding to the minimum
current value (0/4 mA) at the
transmitter output.
Display format from A3.
If two factory settings are given,
then the left-hand one is valid for
CUS 41, the right-hand one for
CUS 31.
O2 (2)
Current output is
simulated
lin
= linear (1)
sim
=
simulation
(2)
Tab = Table (3)
The simulation is terminated by
selecting (1) or (3).
See O2 (1), O2 (3) for other
characteristics.
O221
Enter simulation
value
Current value
0.00 ... 22.00 mA
The current value entered here is
output through the current output.
O2 (3)
Enter current
output table
(only with Plus
package)
lin
= linear (1)
sim = simulation (2)
Tab
=
Table
(3)
Values may also be added or
changed at a later point in time.
The values entered are auto-
matically sorted in ascending
order by current value The
display, however, is not in sorted
order. See O2 (1), O2 (2) for
other characteristics.
Factory settings are printed in
bold
face;
base version does not include functions in
italic.
O211
Sel.Range
4-20
O212
0/4 mA
0.0
FNU
O213
20 mA
9999
FNU
O2
Sel.Type
lin
O221
Simulat.
4.00
mA
O2
Sel.Type
table
UM253E5A.CHP
Liquisys M CUM 223 / 253
Instrument configuration
Hauser
27