5.3
Current outputs
The function group CURRENT OUTPUT is used to configure the individual outputs. Either a
linear (O2 (1)) or, in conjunction with the Plus package, a user-defined current output
characteristic (O2 (3)) can be entered. Furthermore, a current output value can be simulated to
check the current outputs (O2 ( 2)).
Coding
Field
Selection or range
Factory setting
Display
Info
B5
Copy data sets
no
1 -> 2
1 -> 3
2 -> 3
3 -> 2
Data set 1 cannot be changed
(factory setting). However, it can
be used as a basic for a
customer calibration data set.
B6
Display measured
value with
reflection
adaptation?
yes
no
With the CUS 31 / 41: displays
the measured value with or
without assembly adaptation
(reflection). Only effective in
FNU, ppm, mg/l ranges).
B7
Enter corrected
temperature of the
temperature
sensor
Current measured
value
–5.0 ... 100.0 °C
This entry can be used to
calibrate the temperature sensor
to an external measurement.
B8
Enter temperature
difference
(offset)
Current offset
–5,0 ... 5,0 °C
The difference between the
temperatures measured and
entered is displayed.
Factory settings are printed in
bold
face;
base version does not include functions in
italic.
Coding
Field
Selection or range
Factory setting
Display
Info
O
Function group
CURRENT
OUTPUT
Initial display in function group
CURRENT OUTPUT.
O1
Selection of
current output
Out1
Out2
Select the operating mode /
characteristic separately for each
output.
O2 (1)
Enter linear
characteristic
lin
=
linear
(1)
sim = simulation (2)
Tab = Table (3)
The characteristic can have a
positive or negative slope.
See O2 (2), O2 (3) for the
simulation and table
characteristics.
no
B5
copy data
yes
B6
refl.adj.
0.0
B7
RealTemp
°C
0.0
B8
Temp.Offs
°C
O
OUTPUT
O1
Sel. Out
Out1
O2
Sel.Type
lin