3.5

Sensor installation and measuring cable connection

Measuring cable connection

The terminal block is located in a separate
connection compartment which is accessible
by opening the plastic cover. Remove the
pre-pressed knock-outs for cable entry.

The sensor is connected via a multiple-core,
special screened cable which is completely
terminated and permanently connected. Use
a junction box, e.g. VBM to extend the
measuring cable. Termination instructions are
supplied with the measuring cables.

Note:

Protect cable ends and terminals
against moisture to prevent
inaccurate measurement!

For further information on cables
and junction boxes refer to chap.
9 Accessories.

Structure and termination of the measuring cable

Measuring cable requirements

Sensor type

Cable

Extension

Turbidity sensor CUS 31, CUS 41

Non-detachable cable
on sensor

VBM box + CYK 8

Maximum cable length

CUS 31 / CUS 41

max. 200 m using CYK 8

GN

YE

WH

BN

Screen

Com A

Com B

–U

B

+U

B

Sensor

Pin CUM 2x3

S

96

97

88

87

U_AN03E.CDR

+U

B

CYK 8

Fig. 3.13

Structure of the special
measuring cable CYK 8

Installation

Liquisys M CUM 223 / 253

14

Hauser

Содержание Liquisys M CUM 223

Страница 1: ...1500274 Valid from software version 2 20 Liquisys M CUM 223 253 Transmitter for Turbidity and Solids Content ENDRESS HAUSER LIQUISYS S ENDRESS HAUSER LIQUISYS S Hauser Endress The Power of Know How Op...

Страница 2: ...in this chapter Installation You wish to operate or reconfigure the instrument The operating concept is explained in these chapters Operation Instrument configuration Interfaces Need help with proble...

Страница 3: ...ion 18 4 5 Operating concept 19 5 Instrument configuration 5 1 Start up 24 5 2 System configuration 24 5 3 Current outputs 26 5 4 Monitoring functions 28 5 5 Relay contact configuration 31 5 6 Concent...

Страница 4: ...ined until the matter has been settled Check that the delivery is complete and agrees with the shipping documents and your order refer to nameplate for type and version Keep the original packaging mat...

Страница 5: ...ratur output 1 Ausgang 1 output 2 Ausgang 2 mains Netz measuring range Messbereich temperature Temperatur turbidity Tr bung FNU ppm g l 5 70 C 10 55 C 0 4 20 mA 230 VAC 123456 7 5 VA IP 54 IP 30 50 60...

Страница 6: ...ion start up operation Warning This device may only be installed connected electrically commissioned operated and serviced by properly trained personnel authorized by the system operator The personnel...

Страница 7: ...cording to the applicable European standards for industrial applications with regard to electromagnetic compatibility It is carefully protected against electromagnetic interference by constructive des...

Страница 8: ...with integrated temperature sensor and a permanently attachedcable An installation assembly Optional Extension cable CYK 8 Junction box VBM or RM ENDRESS HAUSER LIQUISYS S CAL REL REL1 REL1 ALARM REL...

Страница 9: ...170 247 115 6 157 154 70 70 Pg 13 5 Pg 7 Pg 16 Fig 3 2 Dimensions of Liquisys M CUM 253 FELD2 CDR 1 2 3 4 Fig 3 3 Inside of housing of Liquisys M CUM 253 1 Removable electronics box 2 Partition plate...

Страница 10: ...nting variants Note If installed outdoors weather protection cover required Weather protection cover CYY 101 Weather protection cover for outdoors installation to be mounted on field instrument materi...

Страница 11: ...SS 304 Order no 50086842 Universal mounting post CYY 102 Square tube for mounting of measuring transmitters material stainless steel SS 304 Order no CYY 102 A MONT5 CDR 60 90 70 6 70 90 M6 90 60 x 60...

Страница 12: ...Fig 3 7 Pipe mounting of Liquisys M field instrument MONT2 CDR Fig 3 8 Liquisys M field instrument left Wall mounting right Mounting with universal post and weather protection cover Installation Liqui...

Страница 13: ...Fig 3 9 The required overall installation depth is approx 165 mm MONT9 CDR 92 0 5 92 0 5 90 57 ap prox 25 max 45 96 92 1 2 3 139 6 3 Fig 3 9 Attachment of panel mounted device 1 Wall of control cabin...

Страница 14: ...ng SELV source in accordance with dir IEC 1010 1 Annex H 47 48 49 57 58 59 L L L N 41 42 43 15 V 85 86 10 50 V 81 93 82 94 10 50 V TU mA 31 32 mA 33 34 51 52 53 54 55 56 97 88 87 96 S GN WH BN YE ANU0...

Страница 15: ...5V 10mA GN BN YE WH A REL 4 SENSOR REL 3 REL 2 REL 1 131096 4E CUM 253 Digital 2 Digital 1 87 88 NC Mains Hilfsenergie L1 N AC DC Contacts max 2A AC 250V 500VA DC 30V 60W 131096E CDR Fig 3 11 Connecti...

Страница 16: ...ns are supplied with the measuring cables Note Protect cable ends and terminals against moisture to prevent inaccurate measurement For further information on cables and junction boxes refer to chap 9...

Страница 17: ...onnection examples S 87 88 96 97 U2X3AN CDR BN WH GN YE Sensor Cable CYK 8 Fig 3 14 Connection of turbidity sensors CUS 31 and CUS 41 UM253E03 CHP Liquisys M CUM 223 253 Installation Endress Hauser 15...

Страница 18: ...ys 1 and 2 LED green measured value within permissible limits relay inactive LED red measured value outside permissible limits relay active Alarm indication e g for continuous limit violation temperat...

Страница 19: ...values 1 Temperature display in F 2 Hiding the temperature display 3 Measured value display in FNU 4 Back to basic setting Repeatedly pressing the MINUS key displays the errors 1 The current errors a...

Страница 20: ...operation the relays are controlled by the transmitter REL key The REL key is used to select one of the relays present in the instrument Switch to manual mode The instrument is switched to manual mod...

Страница 21: ...is then reset Code SET MOD CDR Setup mode Access to all configuration settings Measuring mode Standard mode of operation display of current measured values The function groups are selected with the PL...

Страница 22: ...with the PLUS and MINUS keys The ENTER key is used to move from one function to the next within a function group The PLUS and MINUS keys are used for option selection and editing Selections must be c...

Страница 23: ...contacts will go to their normal positions An active hold has priority over all other automatic functions With every hold the I component of the controller is set to zero A possibly accumulated alarm...

Страница 24: ...e of this coding is given in Fig 5 2 The first column indicates the function group as a letter see group designations The functions in the individual groups are counted from the top to the bottom and...

Страница 25: ...20 mA Current output 1 measured value for 4 mA signal current 0 FNU Current output 1 measured value for 20 mA signal current 9999 FNU Current output 2 temperature value at 4 mA signal current 5 0 C C...

Страница 26: ...groups are to be configured to avoid measuring errors or failure to measure at all 5 2 1 Setup 1 Coding Field Selection or range Factory setting Display Info A Function group SETUP 1 Initial display...

Страница 27: ...or range Factory setting Display Info B Function group SETUP 2 Initial display in function group SETUP 2 B1 Switch wiper controller on or off off on auto If auto is selected the wiper is operated in c...

Страница 28: ...r Current measured value 5 0 100 0 C This entry can be used to calibrate the temperature sensor to an external measurement B8 Enter temperature difference offset Current offset 5 0 5 0 C The differenc...

Страница 29: ...one is valid for CUS 41 the right hand one for CUS 31 O2 2 Current output is simulated lin linear 1 sim simulation 2 Tab Table 3 The simulation is terminated by selecting 1 or 3 See O2 1 O2 3 for oth...

Страница 30: ...many times as correspond to the value in O232 Finished appears as the last step After confirmation the system jumps to O236 O234 Enter x value measured value 0 0 FNU 0 0 ppm 0 0 mg l 0 0 g l 0 0 kg l...

Страница 31: ...e fault numbers For individual error numbers please refer to the table in chap 7 The factory settings remain in effect for all errors not edited F6 Make alarm contact effective for selected errors yes...

Страница 32: ...display returns to F If next is selected the display returns to F5 Factory settings are printed in bold face base version does not include functions in italic LIVE CDR Constant measuring signal alarm...

Страница 33: ...ion the relay contact is closed at time t2 when the switch on point has been exceeded t1 and the pickup delay t2 t1 has expired When the alarm threshold t3 is reached and the alarm delay t4 t3 also ha...

Страница 34: ...um control loop stability see table For optimisation reduce the controller gain Kp until the control variable is slightly overswung Increase Kp and then change Tn so that the shortest possible recover...

Страница 35: ...on The greater the calculated control output the higher the switching frequency of the contact The maximum switching frequency 1 T can be adjusted between 60 and 180 min 1 The ON period tON is constan...

Страница 36: ...tacts see chap 5 5 4 Note The timer and Chemoclean do not work independently of each other Whilst one of the functions is active the other cannot be started X X X X w w w w 100 100 Y Y REGEL03 CDR dir...

Страница 37: ...ng cycle Coding Field Selection or range Factory setting Display Info R Function group RELAY Relay contacts can be selected and adjusted R1 Selection of contact to be configured Rel1 Rel2 Rel3 Rel4 Re...

Страница 38: ...nstantly required hysteresis see Fig 5 5 If two factory settings are given then the left hand one is valid for CUS 41 the right hand one for CUS 31 R214 Enter pickup delay 0 s 0 2000 s R215 Enter drop...

Страница 39: ...d see Fig 5 5 R224 Enter pickup delay 0 s 0 2000 s R225 Enter dropout delay 0 s 0 2000 s R226 Enter alarm threshold as an absolute value 150 0 C 20 0 150 0 C When the alarm threshold is exceeded under...

Страница 40: ...See chap 5 5 2 R236 Selection of controller characteristic dir direct inv inverted Setting may or may not be required depending on control deviation up or down deviation see chap 5 5 2 R237 Selection...

Страница 41: ...are activated for the period of time specified here R243 Enter pause time 360 min 1 7200 min The pause time is the time between two cleaning cycles see chap 5 5 4 R244 Enter minimum pause time 120 min...

Страница 42: ...is repeated R257 Enter pause time 360 min 1 7200 min The pause time is the time between two cleaning cycles R258 Enter minimum pause time 120 min 1 R 257 The minimum pause time prevents continuous cl...

Страница 43: ...Info K Function group CONCENTRATION Four different concentration curves can be entered into this function group K1 Selection of concentration curve which will be the basis of the display value calcul...

Страница 44: ...previous settings are kept or back to measurement mode this makes the table invalid Factory settings are printed in bold face base version does not include functions in italic Coding Field Selection o...

Страница 45: ...of an incorrect code returns you to the measurement menu The number is edited with the PLUS or MINUS key and confirmed with the ENTER key S7 Order number is displayed When you upgrade the device S8 Se...

Страница 46: ...E122 E132 E142 Hardware version is displayed This field cannot be edited E113 E123 E133 E143 Serial number is displayed This field cannot be edited E114 E124 E134 E144 Module name is displayed This fi...

Страница 47: ...RFACE I1 Enter address Address HART 0 15 or PROFIBUS 1 126 For communication only The factory settings are printed in bold functions printed in italics are not available in the basic version I INTERFA...

Страница 48: ...253 It is wise to undertake calibration in the turbidity or solids concentration range in which you want to make measurements To simplify the three point calibration we suggest to produce three sample...

Страница 49: ...The calibration can be transferred nonetheless If the calculated results lie outside the permitted limits then the calibration error is displayed E045 The calibration is not transferred Management of...

Страница 50: ...e instructions on the sheet enclosed with the transmitter You will then obtain a 0 10 V current output Factory settings Note Calibration is carried out in the measuring range the transmitter switches...

Страница 51: ...lue from last calibration Immerse sensor in the calibration solution sample 2 Immerse the sensor so that there is sufficient distance to the tank wall no reflection C112 Enter concentra tion of second...

Страница 52: ...bration solution Current value from C113 entire measuring range In case of an unknown probe concentration but with a definite dilution 1 10 1 3 1 the laboratory value is to be entered C122 Calibration...

Страница 53: ...om C113 C133 1 1 C132 C134 Calibration status is displayed o k E xxx C135 Store calibration results yes no new If C134 E xxx then only no or new Exception calibration warning E 84 If new return to C I...

Страница 54: ...Calibration data 6 For FNU Adapter C164 C165 For ppm mg l up to 500 adapter C164 C165 then adapter C166 For g l adapter C166 An existing basic calibration three point is corrected by the single point...

Страница 55: ...Calibration point 3 is displayed Comparison value Deviation relative to standard sensor 100 C164 Slope 1 is displayed Current value Slope of characteristic 1 of the sensor C164 Slope 2 is displayed Cu...

Страница 56: ...5 14 Offset Coding Field Selection or range Factory setting Display Info V Function group OFFSET V1 Enter absolute value Current measured value V2 Enter offset Current offset V3 Calibration status is...

Страница 57: ...For instruments equipped with a communication interface please refer to separate operating instructions BA 208C 07 en HART or BA 209C 07 en PROFIBUS UM253E06 CHP Liquisys M CUM 223 253 Interfaces Endr...

Страница 58: ...value 0 0 and error message E008 Sensor and display in A4 Sensor sensor cable defective Test with new or with different functional sensor CUS 31 or 41 sensor either type is suitable for rough functio...

Страница 59: ...ontactor does not work Controller switched off Activate controller See chap 5 5 or fields R2xx Controller in Manual Off mode Choose Auto or Manual On mode Keyboard REL key Pickup delay setting too lon...

Страница 60: ...nstructions BA 208C HART Field communication with Liquisys M CxM 223 253 HART interface missing Instrument not registered with HART server Load 230 HART receiver e g FXA 191 not connected via load Inc...

Страница 61: ...tion defective Check sensor and sensor connection E H Sevice yes no no E026 Wiper error Check wiper and test function using manual control if necessary yes no no E045 Calibration aborted Repeat calibr...

Страница 62: ...n yes Check relay configuration no no E106 Download yes Wait for download to end no no E116 Download error Repeat download no no E152 PCS alarm Check sensor and connection maintain or replace no no no...

Страница 63: ...p you diagnose problems and provides information about spare parts required Note on the required spare parts You can find information concerning the exact spare part designations and the installation...

Страница 64: ...le required If simulation value is correct check current loop for load and shunts Load excessive Shunt short circuit to frame in current loop Incorrect mode of operation Check whether 0 20 mA or 4 20...

Страница 65: ...R Pos 10 20 Power supply unit LSGA LSGD End frame latches Pos 320 Back plate Pos 310 Side wall for instruments without relay module Pos 40 90 Controller LSCH x LSCP Pos 330 Housing assembly Pos 430 Pu...

Страница 66: ...ning it counterclockwise Remove the complete electronics block from the housing The modules are plugged together mechanically and can be easily separated Pull the controller display module forwards Ge...

Страница 67: ...mperature 51501209 310 Side panel Kit with 10 side panels 51502124 340 End frame PROFIBUS DP End frame with D Submin plug in connector 51502513 330 400 Housing module Housing with front membrane Senso...

Страница 68: ...CP Pos 370 Front membrane Pos 310 320 Ribbon cables Pos 270 Transmitter module MKT1 Pos 10 20 Power supply module LSGA LSGD Pos 340 Docking module LSK1 Pos 290 300 Relay module LSR1 x Pos 450c Torx sc...

Страница 69: ...the module s Dismantling the controller Pos 40 Loosen screw Pos 450b in the display lid Expansion of the docking module Pos 340 Loosen screws Pos 450c in the bottom of the housing and remove entire m...

Страница 70: ...current outputs HART 51501231 80 Controller LSCP PA PROFIBUS PA no current output 51501232 90 Controller LSCP DP PROFIBUS DP no current output 51502499 270 Turbidity transmitter MKT1 Turbidity temper...

Страница 71: ...p CHECK Code P or the Chemoclean function Check the Plus package release e g by opening the function group CHECK Code P or the Chemoclean function Restore the user settings of the instrument 8 4 Spare...

Страница 72: ...lis 5 Ester Hydrocarbons Ketone Household cleaners Caution Do not use any concentrated mineral acids or alkalis benzyl alcohol methylene chloride or high pressure water vapour at over 100 C 8 6 2 Turb...

Страница 73: ...07 en Sensor fixing bracket RM Operating manual BA 205C 07 en Flow Flow assembly FlowFit W CUA 250 Operating manual BA 096C 07 en Welding socket with adapter CUA 120 B Retractable assembly ProbFit CU...

Страница 74: ...version NPT adapter set Order no Connection accessories Extension cable CYK 8 metre goods Unterminated measuring cable for turbidity sensors CUS 31 and CUS 41 Order no 50089633 Extension cable CYK 8 5...

Страница 75: ...l 0 0 200 0 Solids concentration offset range 99 99 FNU 99 99 ppm 99 9 g l 99 9 Temperature measurement Temperature sensor NTC 30 k at 25 C Sensor measuring range 5 0 70 0 C Temperature offset range 5...

Страница 76: ...h ohmic load cos 0 4 max 2 A Switching voltage max 250 V AC 30 V DC Switching power with ohmic load cos 1 max 500 VA AC 60 W DC Switching power with inductive load cos 0 4 max 500 VA AC 60 W DC Limit...

Страница 77: ...ative humidity nominal operating conditions 10 95 non condensing Ingress protection of panel mounted instrument IP 54 front panel IP 30 enclosure Ingress protection of field instrument IP 65 Pollution...

Страница 78: ...C162 Display of cal point 3 Comparative value Display of slope 1 current value C163 C164 Display of slope 2 current value C165 MEAS VALUE DISPLAY with TEMPERATURE DISPLAY in C Measured value display i...

Страница 79: ...ent for previously set error yes no F7 F9 Select next error or return to menu R next next error O235 Entry of y value current value 0 0 20 0 mA 4 00 mA O236 Table status o k no yes Entry of correct pr...

Страница 80: ...for calibration of display value Curve 4 1 K1 Table option selection edit read K3 Set number of value pairs 1 10 1 K4 Value pair selection 1 number of value pairs in 1 K4 K5 Entry of turbidity value e...

Страница 81: ...of alarm threshold as an absolute value entire measuring range 9999 FNU 9999 ppm mg l 300 0 g l 200 0 R216 Entry of post rinse time 0 999 s 20 s R255 Number of repeat cycles 0 5 0 R256 Set interval be...

Страница 82: ...enance CUM 253 66 Corrective maintenance of measuring system 70 CURRENT OUTPUT O 22 26 Current output table 27 Current outputs 26 CYK 8 14 D D component 38 Damage 2 Declaration of conformity 5 Definit...

Страница 83: ...r interface 16 Optoscope 69 Order code 3 43 P P controller 32 P ID controller 31 32 P Function group 30 Packaging 2 Panel mounted instrument 11 Partition plate 7 Pause time 34 39 40 PCS alarm 30 PD co...

Страница 84: ...up menu 24 Stch on 37 Storage 2 Structure of measuring cable 14 SW upgrade 43 Switch off point of contact 36 Switch off temperature 37 Switch on point of contact 36 Symbols 2 Symbols used 2 System con...

Страница 85: ...f health o biological hazardous o inflammable o safe Please mark the appropriate warning hints 5HDVRQ IRU UHWXUQ BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB BBBBBBBBBBB...

Страница 86: ...981 Brazil Samson Endress Hauser Ltda Sao Paulo Tel 55 11 50313455 Fax 11 50313067 Canada Endress Hauser Ltd Burlington Ontario Tel 1 905 6819292 Fax 905 6819444 Chile DIN Instrumentos Ltda Santiago T...