Liquisys M CLM223/253
Technical data
Hauser
99
11
Technical data
11.1
Input
Measured variables
Conductivity, resistivity, temperature
Measuring range
Cable specification
Cell constant
Temperature sensors
Pt 100, Pt 1000, NTC 30K
Measuring frequency
Binary inputs
Current input
4 to 20 mA, galvanically separated
Load: 260
at 20 mA (voltage drop 5.2 V)
Conductivity (conductive):
0 to 600 mS/cm (uncompensated)
Conductivity (inductive):
0 to 2000 mS/cm (uncompensated)
Resistivity:
0 to 200 M
cm
Concentration:
0 to 9999 (%, ppm, mg/l, TDS)
Temperature:
-35 to +250 °C (-31 to +482 °F)
Cable length (conductive):
conductivity: max. 100 m (328 ft) (CYK71)
resistivity: max 15 m (49 ft) (CYK71)
Cable length (inductive):
max 55 m (180 ft) (CLK5)
Cable resistance CYK71:
165
/km (conductivity measurement)
Cable length (conductive):
conductivity: max. 100 m (328 ft) (OYK71)
resistivity: max 15 m (49 ft) (OYK71)
Cable length (inductive):
max 55 m (180 ft) (OLK5)
Cable resistance OYK71:
165
/km (conductivity measurement)
Adjustable cell constant:
k = 0.0025 to 99.99 cm
-1
Conductivity, resistivity (conductive): 170 Hz to 2 kHz
Conductivity (inductive):
2 kHz
Voltage:
10 to 50 V
Power consumption:
max. 10 mA
Содержание Liquisys M CLM223/253
Страница 112: ...www addresses endress com 71212629...