Appendix
Liquisys M CLM223/253
106
Hauser
12
Appendix
Operating matrix
C07-CLM2x3xx-13-06-00-en-001.eps
Entry of
value of
calibration solution
0.00 ... 20.00 %/K
a
2.10 %/K
Calibration temperature
entry (MTC)
–35.0 ... +250.0 C
25.0 C
°
°
Entry of correct
conductivity value of
calibration solution
0.0 µS/cm ... 9999 mS/cm
.
Current meas. value C133
Display of calculated
installation factor
0.10 ... 5.0
1.0
Calibration status
is displayed
o.k.;
E---
Function group
CALIBRATION
C
Calibration
InstF
= Installation factor
Cellc
= Cell constant
AirS = Airset
C1 (3)
C1 (2)
C1 (1)
Residual coupling
Start calibration
Current meas. value
Display of residual coupling
(Airset)
0.0 µS
C112
Calibration status
is displayed
o.k.
E---
.
C113
Store calibration
results
no; new
yes;
C114
C111
C132
C131
C134
C135
Function group
SETUP 1
MEAS. VALUE DISPLAY
with
TEMPERATURE DISPLAY
in °C
A
Selection of operation mode
c
ind = inductive
MOhm = resistance
conc = concentration
ond = conductive
A1
Selection of unit
displayed
µS/cm; mS/cm; S/cm;
µS/m; mS/m; S/m
auto ; k
cm; M
cm; k
auto;
W
W×
W×
W×
m
(omitted if A1 = conc)
Uncompensated
measured value
is displayed
A4
Display format selection
(if A1 = conc)
; X.xxx; XXX.x; XXXX
XX.xx
Temperature display
suppressed
A3
Selection of unit
displayed
; mg/l; %; TDS; none
(% only if A1 = conc)
ppm
Temperature display
in °F
A2
Other errors are displayed
(up to 10 errors)
1st error is displayed
(if present)
Entry of cell constant
cond / ind / MOhm
0.0025 ... 99.99 1/cm for
cond; ind; MOhm
1.000 / 1.98 / 0.01 1/cm
A5
Entry of cable
resistance
(if A1 = cond)
0.00
0.00
ŸW
..
W
. 99.99
A6
Function group
SETUP 2
B
Selection of temperature
measurement
Pt1k (= Pt 1000)
NTC30 (= NTC 30 k
Pt100
W
)
fixed
B1
Selection of temperature
compensation type
.
none
lin = linear
Tab = table
NaCI = common salt
Pure = pure water NaCl
B2
Entry of
value
(if B2 = linear)
0.00 ... 20.00 %/K
a
2.10 %/K
B3
Entry of correct
process temperature
(if B1 = fixed)
25.0 °C
–35.0 °C ... +250.0 °C
B4
Temperature sensor
calibration (omitted
if B1 = fixed)
–35.0 ... +250.0
Display of actual value
°
C
B5
Enter temperature
difference (omitted
if B1 = fixed)
Current offset
–5.0 ... 5.0 °C
B6
Calibration temperature
entry
(if B1 = fixed)
–35.0 ... +250.0 C
25.0 C
.
°
°
C121
Entry of a
value of
calibration solution
0.00 ... 20.00 %/K
a
2.10 %/K
.
C122
Entry of correct
conductivity value of
calibration solution
.
C
urrent meas. value
0.0 mS/cm ... 9999 mS/cm
C123
Display of calculated
cell constant
0.0025 ... 99.99 1/cm
C124
Calibration status
is displayed
o.k.;
E---
C125
Edit mode:
Code 22
Read mode:
an
y code
Edit mode:
Code 22
Read mod:
an
y code
Select alarm
delay unit
; min
s
F2
Alarm delay
0 s ... 2000 s (min)
(depends on F2)
0 s (min)
F3
Error current
setting
2.4 mA
22 mA
F4
Error number
selection
1 ... 255
1
F5
Set alarm contact
to be effective
; no
yes
F6
Function group
ALARM
F
Select contact type
Fleet = fleeting contact
Stead = steady contact;
F1
Function group
CHECK
Switch polarisation
detection on or off
; on
off
Set alarm threshold
Low: High; Lo+Hi;
Lo!; Hi!; LoHi!
Off;
Function group
CURRENT OUTPUT
O
Current output selection
; Out2
Out1
O1
O2
Characteristic selection
table
sim = simulation
lin = linear
O3 (1)
O3 (2)
Simulation value entry
0 ... 22.00 mA
current value
O321
O3 (3)
Table option selection
edit
read
O331
Entry of number of
value pairs in table
1 ... 10
1
O332
Selection of
value pair in table
1 ... number of value pairs
asign
1
O333
Current range selection
0–20 mA
4–20 mA;
O311
O312
Entry of 20 mA value
entire measuring range
2000 mS/cm / 500 k ·cm /
9999 % / 150.0 °C
W
O313
Entry of 0/4 mA value
entire measuring range
0 µS/cm / 0 k ·cm /
0 % / 0 °C
W
Ÿ
Measured value display
Current output in %
Measured value display
Current output in mA
P
P1
P2
P3
P4
P5
P6
Enter alarm delay
0 … 2000 s (min)
0 s (min)
Set lower
alarm threshold
0 … 9999 mS/cm
0 µS/cm
Set upper
alarm threshold
0 … 9999 mS/cm
9999 µS/cm
Z
Z1
Z2
Z3
Z4
Z5
Z6
Z
Cont. switch-off
by current input
Input
Off;
Switch-off direction
for current input
High
Low;
Feedforward control
to PID controller
lin = linear
Off;
Delay for cont.
switch-off current input
0 ... 2000 s
0 s
Delay for cont.
switch-on current input
0 ... 2000 s
0 s
Switch-off limit value
for current input
0 ... 100%
50%
Function group
CURRENT INPUT
Select process
monitoring
AC; CC; AC+CC
AC!; CC!; ACCC!
Off;
Select measured variable
for 2nd current output
°C; mS/cm; Contr
PureH = pure water HCl