Liquisys M CLM223/253
Commissioning
Hauser
43
The values for the sample characteristic (
å
27) are entered in the following table. The
distance per mA can be calculated from
signal /
mA.
First enter the desired current output configuration into the following blank table with a
pencil. Calculate the resulting signal distance per mA to observe the necessary minimum
slope. Then enter the values in the device.
Current output 1
Current output 2
Value pair
[mS/cm]
[%] [°C]
Current
[mA]
Distance
per mA
[mS/cm]
[%] [°C]
Current
[mA]
Distance
per mA
1
500
4
2
1000
16
41.66
3
2000
20
250
Current output 1
Current output 2
Value pair
[mS/cm]
[%] [°C]
Current
[mA]
Distance
per mA
[mS/cm]
[%] [°C]
Current
[mA]
Distance
per mA
1
2
3
4
5
6
7
8
9