Liquisys M CLM223/253
Diagnostics and troubleshooting
Hauser
79
7.4
Instrument specific errors
The following table helps you during the diagnosis and points to any spare parts required.
Depending on the degree of difficulty and the measuring equipment present, diagnosis is
carried out by:
• Trained operator personnel
• The user's trained electrical technicians
• Company responsible for system installation/operation
• Hauser Service
Information on the exact spare part designations and on how to install these parts can be
found in the "Spare parts" section.
Error
Possible cause
Tests and/or remedial measures
Execution, tools, spare parts
Device cannot be
operated, display value
9999
Operation locked
Press CAL and MINUS keys
simultaneously.
See "Function of keys" section.
Display dark, no light-
emitting diode active
No line voltage
Check whether line voltage is present.
Electrical technician/e.g. multimeter
Supply voltage wrong/too low
Compare actual line voltage and
nameplate data.
User
(data for energy supply company or
multimeter)
Connection faulty
Terminal not tightened;
insulation jammed;
wrong terminals used.
Electrical technician
Device fuse defective
Compare line voltage and the nameplate
data and replace fuse.
Electrical technician/suitable fuse;
see drawing in "Spare parts" section.
Power unit defective
Replace power unit, note variant.
On-site diagnosis by Hauser
Service, test module necessary
Central module defective
Replace central module, note variant.
On-site diagnosis by Hauser
Service, test module necessary
CLM253: ribbon cable item 310
loose or defective
Check ribbon cable, renew if necessary.
See "Spare parts" section.
Display dark, light-
emitting diode active
Central module defective
(module: LSCH/LSCP)
Renew central module, note variant.
On-site diagnosis by Hauser
Service, test module necessary
Display is on but
– No change in display
and/or
– Device cannot be
operated
Device or module in device not
correctly mounted
CLM223: reinstall insert.
CLM253: remount display module.
Perform with the aid of the installation
drawings in the "Spare parts" section.
Operating system in unpermitted
mode
Switch device off and then on again.
Poss. EMC problem: if this persists, check
the installation or have it checked by
Hauser Service.
Device gets hot
Voltage wrong/too high
Compare line voltage and nameplate
data.
User, electrical technician
Power unit defective
Replace power unit.
Diagnosis only by Hauser
Service
Incorrect meas. cond/M
and/or temperature
Transmitter module defective
(module: MKIC), please first carry
out tests and take measures as per
the "Process errors without
messages" section.
Measuring input test:
– Connect resistor in place of
conductivity sensor
– Resistance 100
at terminals 11/
12 + 13 = display 0 °C
If test negative: replace module (note
variant).
Perform with the aid of the exploded
drawings in the "Spare parts" section.
Current output, current
value incorrect
Adjustment not correct
Check with installed current simulation,
connect mA meter directly to current
output.
If simulation value incorrect: adjustment
in factory or new module LSCxx required.
If simulation value correct: check current
loop for load and shunts.
Load too big
Shunt/short to ground in current
loop
Incorrect mode of operation
Check whether 0–20 mA or 4–20 mA is
selected.
No current output signal
Current output stage defective
(module LSCH/LSCP)
Check with installed current simulation,
connect mA meter directly to current
output.
If test negative:
Renew central module LSCH/LSCP (note
variant).