Liquisys M CLM223/253
Commissioning
Hauser
55
EP-PW function:
For "Purified Water" (PW) according to the standard of EP the transmitter
uses the temperature depending limit values of the following table:
The measurement is performed as follows:
• The transmitter determines the uncompensated conductivity and the water temperature .
• The transmitter determines the conductivity limit value by interpolation of two table
values.
• If the measured value exceeds the table value an alarm (E151) will be triggered.
Prealarm:
Also available is a pre-alarm with a user defined switch-on point (e.g. 80 % of the
USP/EP value). This gives the user a signal for the in-time regeneration of the system. Enter
the value in field R262 or R272.
Pay attention to the following:
• To use the USP and EP functions the transmitter has to be equipped with a relay board and
the Plus Package.
• For alarm output, activate the fault-signalling contact or the error current in field F5 - F7
(error code E151 and E153).
• The switch-off point of the pre-alarm is 1 % below of the switch-on point (R262 / R272),
related to the main limit value.
• The transmitter uses uncompensated values for the USP and EP functions even in case
temperature compensated values are displayed.
• At temperatures above 100 °C (212 °F) the transmitter uses the limit value of 100 °C
(212 °F).
Temperature [°C/°F]
Conductivity [
μ
S/cm]
Temperature [°C/°F]
Conductivity [
μ
S/cm]
0/32
2.4
60/140
8.1
10/50
3.6
70/158
9.1
20/68
4.3
75/167
9.7
25/77
5.1
80/176
9.7
30/86
5.4
90/194
9.7
40/104
6.5
100/212
10.2
50/122
7.1