Liquisys M CLM223/253
Operability
Hauser
29
Menu structure
The configuration and calibration functions are arranged in function groups.
• In setup mode, select a function group with the PLUS and MINUS keys.
• In the function group itself, switch from function to function with the ENTER key.
• Within the function, select the desired option with the PLUS and MINUS keys or edit the
settings with these keys. Then confirm with the ENTER key and continue.
• Press the PLUS and MINUS keys simultaneously (Escape function) to exit programming
(return to the main menu).
• Press the PLUS and MINUS simultaneously keys again to switch to the measuring mode.
If a modified setting is not confirmed with ENTER, the old setting is retained.
An overview of the menu structure is provided in the Appendix to these Operating
Instructions.
C07-CxM2x3xx-19-06-00-xx-002.eps
Fig. 19:
Diagram of the menu structure
1
Functions (parameters selected, numbers entered)
2
Function groups, scroll backwards and forwards with the PLUS and MINUS keys
3
Switch from function to function with the ENTER key
Hold function: "freezing" of the outputs
During setup and calibration, the current output can be "frozen". It constantly retains its
current status. "HOLD" appears on the display. If the controller actuating variable (steady
control 4 to 20 mA) is output via current output 2, it is set to 0/4 mA in Hold.
Pay attention to the following:
• Hold settings can be found in the "Service" section.
• During Hold, all contacts will go to their normal positions.
• An active Hold has priority over all other functions.
• With every Hold, the I-component of the controller is set to zero.
• Any alarm delay is reset to "0".
• This function can also be activated externally via the Hold input (see Wiring diagram;
binary input 1).
• The manual Hold (field S3) remains active even after a power failure.