Liquiphant FTL62 Density
Technical data
Hauser
31
15.3.9 Mechanical load
Support the device in the event of severe dynamic load. Maximum lateral loading capacity
of the pipe extensions and sensors: 75 Nm (55 lbf ft).
A
For more details, see the "Supporting the device" section.
15.3.10 Pollution degree
Pollution degree 2
15.3.11 Electromagnetic compatibility
Electromagnetic compatibility as per EN 61326 series and NAMUR recommendation EMC
(NE21).
15.4 Process
15.4.1 Process temperature range
0 to 80 °C (32 to 176 °F)
15.4.2 Thermal shock
≤ 120 K/s
15.4.3 Process pressure range
–1 to +25 bar (–14.5 to +362.5 psi)
L
WARNING
The maximum pressure for the device depends on the lowest-rated element, with
regard to pressure, of the selected component. This means that it is necessary to pay
attention to the process connection as well as the sensor.
‣
Pressure specifications, see the associated Technical Information.
‣
Only operate the device within the specified limits!
‣
The Pressure Equipment Directive (2014/68/EU) uses the abbreviation "PS". The
abbreviation "PS" corresponds to the MWP (maximum working pressure) of the device.
Canadian CRN approval: more details about the maximum pressure values are
available in the download area of the product page under:
www.endress.com
→
Downloads.
15.5 Additional technical data
• Technical information Liquiphant FTL62 Density: TI01553F
• Technical information Liquiphant FTL62: TI01539F
Содержание Liquiphant FTL62 Density
Страница 33: ......
Страница 34: ......
Страница 35: ......