Mounting
Liquiphant FTL62 Density
14
Hauser
D
r
80 mm (3.15 in)
1.0008
82 mm (3.23 in)
1.0006
84 mm (3.31 in)
1.0005
86 mm (3.39 in)
1.0004
88 mm (3.46 in)
1.0003
90 mm (3.54 in)
1.0003
92 mm (3.62 in)
1.0002
94 mm (3.70 in)
1.0002
96 mm (3.78 in)
1.0001
98 mm (3.86 in)
1.0001
100 mm (3.94 in)
1.0001
>100 mm (3.94 in)
1.0000
Installation in pipe
Correction factor "r" as a factor of the internal diameter of pipe "D", for entering in the
Density Computer FML621 or ReadWin2000:
D
r
D
A0039707
<44 mm (1.73 in)
-
44 mm (1.73 in)
1.0225
46 mm (1.81 in)
1.0167
48 mm (1.89 in)
1.0125
50 mm (1.97 in)
1.0096
52 mm (2.05 in)
1.0075
54 mm (2.13 in)
1.0061
56 mm (2.20 in)
1.0051
58 mm (2.28 in)
1.0044
60 mm (2.36 in)
1.0039
62 mm (2.44 in)
1.0035
64 mm (2.52 in)
1.0032
66 mm (2.60 in)
1.0028
68 mm (2.68 in)
1.0025
70 mm (2.76 in)
1.0022
72 mm (2.83 in)
1.0020
74 mm (2.91 in)
1.0017
76 mm (2.99 in)
1.0015
78 mm (3.07 in)
1.0012
80 mm (3.15 in)
1.0009
82 mm (3.23 in)
1.0007
84 mm (3.31 in)
1.0005
86 mm (3.39 in)
1.0004
88 mm (3.46 in)
1.0003
90 mm (3.54 in)
1.0002