Liquiphant FTL62 Density
Electrical connection
Hauser
19
Terminal assignment
The output signal of the density sensor is based on pulse technology. With the aid of this
signal, the fork frequency is continuously forwarded to the Density Computer FML621.
–
+
Pulse
1
2
!
82
182
10
112
110
113
83
183
-
.
-
-
+
+
A, E
B, C, D
1
2
3
4
4
3
1
2
1
3
4
M12
L+ L- PE
B
A
A0036059
14
Connection diagram: Connecting the FEL60D electronic insert to the Density Computer FML621
A
Connection wiring with terminals
B
Connection wiring with M12 plug in housing according to EN61131-2 standard
1
Electronic insert FEL60D
2
Density Computer FML621
3
Slots A, E with extension cards (already included in the basic unit)
4
Slots B, C, D with extension cards (optional)
Supply voltage
U = 24 V
DC
± 15 %, only suitable for connecting to the Density Computer FML621
Power consumption
P < 160 mW
Current consumption
I < 10 mA
Overvoltage protection
Overvoltage category II