Certificates and approvals
CE mark
The product meets the requirements of the harmonized European standards. As such, it complies
with the legal specifications of the EC directives. The manufacturer confirms successful testing of the
product by affixing to it the CE-mark.
Hygiene standard
• EHEDG certification, type EL - CLASS I. Permitted process connections in accordance with EHEDG,
see 'Process connections' section → 29
• 3-A authorization no. 1144, 3-A sanitary standard 74-06. Permitted process connections in
accordance with 3-A, see 'Process connections' section → 29
• ASME BPE 2009, certificate of conformity can be ordered for indicated options
• FDA-compliant
• All product contact surfaces are produced without animal fats (TSE Certificate of Suitability)
Ex approval
Information about currently available Ex versions (ATEX, FM, CSA, etc.) can be supplied by your E+H
Sales Center on request. All explosion protection data are given in separate documentation which is
available upon request.
Other standards and
guidelines
• EN 60079: ATEX certification for hazardous areas
• IEC 60529: Degrees of protection provided by enclosures (IP code)
• IEC 61010-1: Safety requirements for electrical equipment for measurement, control and
laboratory use
• IEC 60751: Industrial platinum resistance thermometers
• ASTM E 1137/E1137M-2008: Standard Specification for Industrial Platinum Resistance
Thermometers
• EN 50281-1-1: Electrical apparatus protected by enclosures
• DIN 43772: Protection tubes
• DIN EN 50446: Terminal heads
• IEC 61326-1: Electromagnetic compatibility (electrical equipment for measurement, control and
laboratory use - EMC requirements)
Parts in contact with the
medium
Parts of the thermometer in contact with the medium comply with the following European
regulations:
• (EC) No. 1935/2004, Article 3, paragraph 1, Articles 5 and 17 on materials and articles intended
to come into contact with food.
• (EC) No. 2023/2006 on good manufacturing practice for materials and articles intended to come
into contact with food.
• (EC) No. 10/2011 on plastic materials and articles intended to come into contact with food.
• Canadian Registration Number (CRN):
The CRN approval is only available for certain options of protection tubes. These will be marked
and shown during the configuration of this device.
Surface purity
Free from oil and grease, optional
Material resistance
Material resistance - including terminal heads - to the following cleaning agents/disinfectants from
the company Ecolab: P3-topax 66, P3-topactive 200, P3-topactive 500 and P3-topactive OKTO as
well as demineralized water.
Material certification
The material certificate 3.1 (according to standard EN 10204) can be requested separately.
Calibration
The "Factory calibration" is carried out according to an internal procedure in a laboratory of Endress
+Hauser accredited by the European Accreditation Organization (EA) to ISO/IEC 17025. A
calibration which is performed according to EA guidelines (SIT/Accredia) or (DKD/DAkkS) may be
requested separately. The calibration is performed on the replaceable insert of the thermometer. In
the case of thermometers without a replaceable insert, the entire thermometer - from the process
connection to the tip of the thermometer - is calibrated.
Содержание iTHERM TM401