Hauser
Field housing T17 for TMT162
Specification
°C
10
0
20
30
40
50
114 (4.49)
114 (4.49)
94 (3.7)
A0034506
• Protection class:
– IP66
– IP68
• Temperature:
– Without display: –40 to +85 °C (–40 to +185 °F)
– With display: –40 to +80 °C (–40 to +80 °F)
• Material: Stainless steel 1.4404/316L for hygienic
applications
• Threaded cable entry: 1x or 2x M20x1.5, G½" or ½" NPT
• Protection armature connection: ½" NPT
• Display rotatable in 90° increments
• Brilliant blue backlit display with ease of visibility in
bright sunshine or pitch darkness
• Separate electronics compartment and connection
compartment
• Gold plated terminals to avoid corrosion and additional
measurement errors
• Weight: 1.25 kg (2.76 lb)
• Available in conjunction with 3-A marked sensors
Cable glands and fieldbus connectors
Type
Suitable for cable entry
Degree of protection
Temperature range
Cable gland, polyamide
½" NPT, ¾" NPT,
M20x1.5 (optionally 2x
cable entry)
IP68
–40 to +100 °C (–40 to +212 °F)
½" NPT, M20x1.5
IP69K
–20 to +95 °C (–4 to +203 °F)
Cable gland for dust ignition-proof area, polyamide
½" NPT, M20x1.5
(optionally 2x cable
entry)
IP68
–20 to +95 °C (–4 to +203 °F)
Fieldbus connector (M12x1 PA, 7/8" PA, FF)
½" NPT, M20x1.5
IP67, NEMA Type 6
–40 to +105 °C (–40 to +221 °F)
Fieldbus connector (M12, 8-pin)
M20x1.5
IP67
–30 to +90 °C (–22 to +194 °F)
Extension neck
Standard version of extension neck, or optionally with the quick-fastening iTHERM QuickNeck.
• Tool-free removal of the insert:
– Saves time/costs on frequently calibrated measuring points
– Wiring mistakes avoided
• IP69K protection class
Содержание iTHERM TM401