Process connections for thermometer installation in pipes with small nominal diameters
1
Varivent
®
- process connection D = 50 mm for DN25 pipes
2
Clamp or micro-clamp
1
2
3
Sensor
Varivent
with
connection
Shaped
gasket
Companion
connection
O-ring
Companion
connection
Gasket
(O-ring)
Welding boss
Leak detection hole
Vessel wall
A0033600-EN
6
Detailed installation instructions for hygiene-compliant installation
1
Varivent
®
process connection for VARINLINE
®
housing
2
Clamp according to ISO 2852
3
Liquiphant-M G1" process connection, horizontal installation
The counterpieces for the process connections and the seals or sealing rings are not included in
the scope of supply for the thermometer. Liquiphant M weld-in adapters with associated seal
kits are available as accessories. → 35
Procedure in case of seal failure indicated by leak detection port:
• Disassembling of the thermometer, validated cleaning procedure of thread and sealing ring
groove
• Replacement of the seal or sealing ring
• CIP after re-assembly
In the case of weld-in connections, exercise the necessary degree of care when performing the
welding work on the process side:
• Suitable welding material
• Flush-welded or with welding radius > 3.2 mm (0.13 in)
• No recesses, folds or gaps
• Honed and polished surface, Ra ≤ 0.76 µm (30 µin)
As a general rule, the thermometers should be installed in such a way that does not impact
their ability to be cleaned (the requirements of the 3-A Sanitary Standard must be observed).
The Varivent
®
and Liquiphant-M weld-in adapter enable flush-mounted installation.