6.4
Configuration
6.4.1
One current output: active/passive
The current output is configured as "active" or "passive" by means of various jumpers on the I/O
board.
"
Caution!
Configuring the current outputs as "active" or "passive" is possible for non-Ex i I/O boards only.
Ex i I/O boards are permanently wired as "active" or "passive". Also refer to table
#
Warning!
Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the power
supply is switched off before you remove the cover of the electronics compartment.
1.
Switch off power supply.
2.
3.
Set the jumpers in accordance with
"
Caution!
Risk of destroying the measuring device. Set the jumpers exactly as shown in the diagram.
Incorrectly set jumpers can cause overcurrents that would destroy either the measuring device or
external devices connected to it.
4.
Installation of the I/O board is the reverse of the removal procedure.
a0001044
Fig. 30:
Configuring the current output (I/O board)
1
Active current output (default)
2
Passive current output
1
2