Proline Promass 80
Operation
Hauser
43
13-5
586
The fluid properties do not allow normal measuring operation.
13-6
587
Extreme process conditions exist.
The measuring system can therefore not be started.
13-7
588
Overdriving of the internal analog to digital converter.
A continuation of the measurement is no longer possible!
14-3
601
Positive zero return active
14-7
611
Simulation current output 1 active
15-0
612
Simulation current output 2 active
15-3
621
Simulation frequency output active
15-7
631
Simulation pulse output active
16-3
641
Simulation status output active
17-7
671
Simulation status input active
18-3
691
Simulation of response to error (outputs) active
18-4
692
Simulation of measured variable active
19-0
700
The process fluid density is outside the upper or lower limit values
19-1
701
The maximum current value for the measuring tube exciter coils has been
reached,
since certain process fluid characteristics are extreme.
19-2
702
Frequency control is not stable, due to inhomogeneous fluid.
19-3
703
Overdriving of the internal analog to digital converter.
A continuation of the measurement is still possible!
19-4
704
19-5
705
The electronics' measuring range will be exceeded. The mass flow is too high.
20-5
731
The zero point adjustment is not possible or has been canceled.
Byte–bit
Error No.
Short error description
ä
56