Technical data
Proline Promass 80
84
Hauser
Example for max. measured error
A0013425
Fig. 40:
Max. measured error in % o.r. (example: Promass H, DN 25)
Flow values (example)
Repeatability
Design fundamentals
85.
o.r. = of reading; 1 g/cc = 1 kg/l; T = medium temperature
Material measuring tube: Zirconium 702/R 60702
• Mass flow and volume flow (liquid): ±0.05% o.r.
• Density (liquid): ±0.00025 g/cc
• Temperature: ±0.25 °C ± 0.0025 · T °C; (±0.5 °F ± 0.0015 · (T–32) °F)
Material measuring tube: Tantalum 2.5W
• Mass flow and volume flow (liquid): ±0.05% o.r.
• Mass flow (gas): ±0.25% o.r.
• Density (liquid): ±0.00025 g/cc
• Temperature: ±0.25 °C ± 0.0025 · T °C; (±0.5 °F ± 0.0015 · (T–32) °F)
Influence of medium temperature
When there is a difference between the temperature for zero point adjustment and the process
temperature, the typical measured error of the sensor is ±0.0002% of the full scale value / °C
(±0.0001% of the full scale value/°F).
0
±0.5
±0.2
±1.0
[%]
0
1
2
4
6
8
10
12
14
16
18
t/h
Turn down
Flow
Maximum measured error
[kg/h] or [l/h]
[lb/min]
[% o.r.]
250 : 1
72
2.646
2.50
100 : 1
180
6.615
1.00
25 : 1
720
26.46
0.25
10 : 1
1800
66.15
0.10
2 : 1
9000
330.75
0.10
Содержание HART Proline Promass 80
Страница 2: ......
Страница 114: ...www endress com worldwide BA00057D 06 EN 13 12 71197479 FM SGML 10 0 ProMoDo ...